Zobrazit minimální záznam

dc.contributor.authorZach, Lukáš
dc.contributor.authorKunčická, Lenka
dc.contributor.authorRůžička, Pavel
dc.contributor.authorBaghirova, Banovsha
dc.date.accessioned2016-04-27T13:47:42Z
dc.date.available2016-04-27T13:47:42Z
dc.date.issued2016
dc.identifier.citationMetalurgija. 2016, vol. 55, no. 3, p. 349-352.cs
dc.identifier.issn0543-5846
dc.identifier.issn1334-2576
dc.identifier.urihttp://hdl.handle.net/10084/111516
dc.description.abstractThe aim of this study is to create and verify a numerical model for a Medin Modular orthopedic knee-joint implant by investigating contact pressure, its distribution and contact surfaces. An experiment using Fuji Prescale pressure sensitive films and a finite element analysis (FEA) using Abaqus software were carried out. The experimental data were evaluated using a special designed program and were compared with the results of the analysis. The designed evaluation program had been constructed on the basis of results obtained from a supplementary calibration experiment. The applicability of the numerical model for the real endoprosthesis behavior prediction was proven on the basis of their good correlation.cs
dc.format.extent189270 bytes
dc.format.mimetypeapplication/pdf
dc.language.isoencs
dc.publisherHrvatsko Metalurško Društvocs
dc.relation.ispartofseriesMetalurgijacs
dc.relation.urihttp://hrcak.srce.hr/file/226316cs
dc.subjectendoprosthesiscs
dc.subjectknee jointcs
dc.subjectnumericalcs
dc.subjectanalysiscs
dc.subjectFEMcs
dc.titleExperimental and numerical analysis of a knee endoprosthesis numerical modelcs
dc.typearticlecs
dc.type.statusPeer-reviewedcs
dc.description.sourceWeb of Sciencecs
dc.description.volume55cs
dc.description.issue3cs
dc.description.lastpage352cs
dc.description.firstpage349cs
dc.identifier.wos000372344500015


Soubory tohoto záznamu

Tento záznam se objevuje v následujících kolekcích

Zobrazit minimální záznam