dc.description.abstract | The purpose of this thesis is to calibrate a high-volume gamma-ray scintillation spectrometer
in energy levels higher than those of natural radionuclides and common calibration standards. It is
achieved by a method called neutron activation analysis, that utilizes slow and 14 MeV neutrons to
bombard elements of H, C, N and O. These activated elements then dexcite with emission of gammarays with energies up to 11 MeV. The calibration on common standards allows us to only calibrate in
the energy range below 3 MeV. The energy calibration in this range does not apply in the higher energy
range as the calibration shifts with great deviations. Total of 36 measurements were performed to
gather data on which the calibration would be carried out. | en |
dc.description.abstract | Cílem této práce je kalibrace velkoobjemového scintilačního spektrometru gama záření v
oblasti energií vyšších, než vyzařují přirozené radionuklidy a běžné kalibrační etalony. Toho je
dosáhnuto metodou zvaná “neutronová aktivační analýza”, která využívá pomalých a 14 MeV neutronů
k bombardování prvků H, C, N a O. Tyto aktivované prvky poté podléhají deexcitaci doprovázená emisí
gama záření s energiemi až 11 MeV. Kalibrace na běžných etalonech nám umožňuje kalibrovat pouze
v oblasti energií do 3 MeV. Energetická kalibrace v této oblasti přestává platit v oblasti vyšších energií,
kde dochází k posunu kalibrace s velkými odchylkami. Dohromady bylo provedeno 36 měření, aby byla
získána data, na kterých by bylo možné kalibraci provést. | cs |
dc.format.extent | 3062252 bytes | |
dc.format.mimetype | application/pdf | |
dc.language.iso | en | |
dc.publisher | Vysoká škola báňská – Technická univerzita Ostrava | cs |
dc.subject | Fast Neutrons | en |
dc.subject | Gamma Spectroscopy | en |
dc.subject | Neutron Generator | en |
dc.subject | Neutron Activation Analysis | en |
dc.subject | Energy
Calibration | en |
dc.subject | Rychlé neutrony | cs |
dc.subject | Gama Spektrometrie | cs |
dc.subject | Neutronový Generátor | cs |
dc.subject | Neutronová Aktivační Analýza | cs |
dc.subject | Energetická Kalibrace | cs |
dc.title | Calibration of a high-volume gamma-ray scintillation spectrometer in the high-energy range | en |
dc.title.alternative | Kalibrace velkoobjemového scintilačního spektrometru gama záření v oblasti vyšších energií | cs |
dc.type | Bakalářská práce | cs |
dc.contributor.referee | Brunclík, Tomáš | |
dc.date.accepted | 2022-05-30 | |
dc.thesis.degree-name | Bc. | |
dc.thesis.degree-level | Bakalářský studijní program | cs |
dc.thesis.degree-grantor | Vysoká škola báňská – Technická univerzita Ostrava. Fakulta elektrotechniky a informatiky | cs |
dc.description.department | 480 - Katedra fyziky | cs |
dc.thesis.degree-program | Aplikovaná fyzika | cs |
dc.description.result | výborně | cs |
dc.identifier.sender | S2724 | |
dc.identifier.thesis | KUC0332_FEI_B0533A110023_2022 | |
dc.rights.access | openAccess | |