dc.contributor.authorMareš, Radim
dc.date.accessioned2011-10-10T07:24:31Z
dc.date.available2011-10-10T07:24:31Z
dc.date.issued2010
dc.identifier.citationSborník vědeckých prací Vysoké školy báňské - Technické univerzity Ostrava. Řada strojní. 2010, roč. 56, č. 3, s. 115-118 : il.cs
dc.identifier.issn1210-0471cs
dc.identifier.urihttp://hdl.handle.net/10084/89157
dc.description.abstractTemperature in a chamber for measuring surface tension of metastable supercooled water by the elevation method is detected by resistance temperature sensors. Two copper blocks of different temperatures placed on two opposite sides of the measuring chamber serve as heat accumulators. This paper deals with the influence of heat radiation of the copper blocks on the sensor depending on its position in the chamber.cs
dc.format.extent231159 bytes
dc.format.mimetypeapplication/pdf
dc.language.isoencs
dc.publisherVysoká škola báňská - Technická univerzita Ostravacs
dc.relation.ispartofseriesSborník vědeckých prací Vysoké školy báňské - Technické univerzity Ostrava. Řada strojnícs
dc.relation.urihttp://transactions.fs.vsb.cz/2010-3/1823.pdf
dc.rights© Vysoká škola báňská - Technická univerzita Ostrava
dc.titleInfluence of heat radiation on a temperature sensor in a measuring chambercs
dc.title.alternativeVliv sálání na teplotní čidlo v měřicí komůrcecs
dc.typearticlecs
dc.description.abstract-enTeplota v komůrce pro měření povrchového napětí metastabilní podchlazené vody elevační metodou je snímána odporovými teplotními čidly. Dva měděné bloky o rozdílných teplotách umístěné na protilehých stranách komůrky slouží jako akumulátory tepla. Článek se zabývá vlivem sálavého tepla měděných bloků na teplotu odporového čidla v závislosti na jeho poloze v komůrce.cs
dc.rights.accessopenAccess