The detection properties comparison for the SnO2 gas sensor in different heating regimes and the step change in concentration

Loading...
Thumbnail Image

Downloads

3

Date issued

Journal Title

Journal ISSN

Volume Title

Publisher

Vysoká škola báňská - Technická univerzita Ostrava

Location

Signature

License

Abstract

Non-periodical heating regimes and also periodical heating regimes were applied for the sensors TGS 813 and SP 11. Ethanol vapour in air was used as a detected substance. It was found that the sensitivity is better in the periodical square-wave heating regime among the tested regimes. The stabilisation of the sensor conductivity, when the concentration abruptly varies its value, takes longer compared to the constant heating regime at maximum heating voltage.

Description

Subject(s)

dynamic response, semiconductor gas sensor, temperature modulation, thermal cycling, tin dioxide

Citation

Advances in electrical and electronic engineering. 2016, vol. 14, no. 5, p. 583-590