Využití detektoru DAVID pro verifikaci IMRT plánu v reálném čase
Loading...
Downloads
4
Date issued
Authors
Journal Title
Journal ISSN
Volume Title
Publisher
Vysoká škola báňská – Technická univerzita Ostrava
Location
Signature
Abstract
Unlike single point verification of classic in-vivo dosimetry, today with help of DAVID system it is possible to verify the radiation plan in the whole width of the irradiated area. The DAVID system is a multiwire ionization chamber, which measures dose below leaves of a collimator. Mostly it is used for IMRT, which is based on field modulation by multileaf collimator.
This thesis researches possibilities of the DAVID system for verification of patients’ radiation plans. The thesis goal is to gather data for planning system, creation of radiation plans and their verification with help of the independent methods. The second verification system was OCTAVIUS Seven29.
The measurement has been performed with the linear accelerator TrueBeam by Varian company at the oncology department of Nemocnice České Budějovice a.s.
Description
Subject(s)
DAVID, IMRT, Gamma analysis, VeriSoft, Linear accelerator, OCTAVIUS Seven29, Radiotherapy, MLC