Experimental and numerical analysis of a knee endoprosthesis numerical model

Loading...
Thumbnail Image

Downloads

5

Date issued

Authors

Zach, Lukáš
Kunčická, Lenka
Růžička, Pavel
Baghirova, Banovsha

Journal Title

Journal ISSN

Volume Title

Publisher

Hrvatsko Metalurško Društvo

Location

Signature

Abstract

The aim of this study is to create and verify a numerical model for a Medin Modular orthopedic knee-joint implant by investigating contact pressure, its distribution and contact surfaces. An experiment using Fuji Prescale pressure sensitive films and a finite element analysis (FEA) using Abaqus software were carried out. The experimental data were evaluated using a special designed program and were compared with the results of the analysis. The designed evaluation program had been constructed on the basis of results obtained from a supplementary calibration experiment. The applicability of the numerical model for the real endoprosthesis behavior prediction was proven on the basis of their good correlation.

Description

Subject(s)

endoprosthesis, knee joint, numerical, analysis, FEM

Citation

Metalurgija. 2016, vol. 55, no. 3, p. 349-352.