dc.contributor.author | Ličev, Lačezar | |
dc.contributor.author | Krumnikl, Michal | |
dc.contributor.author | Škuta, Jaromír | |
dc.contributor.author | Babiuch, Marek | |
dc.contributor.author | Farana, Radim | |
dc.date.accessioned | 2015-01-27T14:35:36Z | |
dc.date.available | 2015-01-27T14:35:36Z | |
dc.date.issued | 2014 | |
dc.identifier.citation | Biotechnology & Biotechnological Equipment. 2014, vol. 28, issue 2, s. 355-359. | cs |
dc.identifier.issn | 1310-2818 | |
dc.identifier.issn | 1314-3530 | |
dc.identifier.uri | http://hdl.handle.net/10084/106362 | |
dc.description.abstract | This paper describes the advances in the development and subsequent testing of an imaging device for three-dimensional ultrasound measurement of atherosclerotic plaque in the carotid artery. The embolization from the atherosclerotic carotid plaque is one of the most common causes of ischemic stroke and, therefore, we consider the measurement of the plaque as extremely important. The paper describes the proposed hardware for enhancing the standard ultrasonic probe to provide a possibility of accurate probe positioning and synchronization with the cardiac activity, allowing the precise plaque measurements that were impossible with the standard equipment. The synchronization signal is derived from the output signal of the patient monitor (electrocardiogram (ECG)), processed by a microcontroller-based system, generating the control commands for the linear motion moving the probe. The controlling algorithm synchronizes the movement with the ECG waveform to obtain clear images not disturbed by the heart activity. | cs |
dc.format.extent | 403458 bytes | |
dc.format.mimetype | application/pdf | |
dc.language.iso | en | cs |
dc.publisher | Taylor & Francis | cs |
dc.relation.ispartofseries | Biotechnology & Biotechnological Equipment | cs |
dc.relation.uri | https://doi.org/10.1080/13102818.2014.910362 | cs |
dc.rights | © 2014 The Author(s). Published by Taylor & Francis. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/3.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. The moral rights of the named author(s) have been asserted. Permission is granted subject to the terms of the License under which the work was published. Please check the License conditions for the work which you wish to reuse. Full and appropriate attribution must be given. This permission does not cover any third party copyrighted material which may appear in the work requested. | cs |
dc.title | Advances in the development of an imaging device for plaque measurement in the area of the carotid artery | cs |
dc.type | article | cs |
dc.identifier.doi | 10.1080/13102818.2014.910362 | |
dc.rights.access | openAccess | |
dc.type.version | publishedVersion | cs |
dc.type.status | Peer-reviewed | cs |
dc.description.source | Web of Science | cs |
dc.description.volume | 28 | cs |
dc.description.issue | 2 | cs |
dc.description.lastpage | 359 | cs |
dc.description.firstpage | 355 | cs |
dc.identifier.wos | 000342313600027 | |