dc.description.abstract | A classical method for estimation of respiratory information from electrocardiogram (ECG), called ECG - derived respiration (EDR), is using flexible electrodes located at standard electrocardiography positions. This work introduces an alternative approach suitable for miniaturized sensors with low inter-electrode separation and electrodes fixed to the sensor encapsulation. Application of amplitude EDR algorithm on single-lead wireless sensor system with optimized electrode positions shows results comparable with standard robust systems. The modified method can be applied in daily physiological monitoring, in sleep studies or implemented in smart clothes when standard respiration techniques are not suitable. | cs |
dc.language.iso | en | cs |
dc.publisher | Vysoká škola báňská - Technická univerzita Ostrava | cs |
dc.relation.ispartofseries | Advances in electrical and electronic engineering | cs |
dc.relation.uri | http://dx.doi.org/10.15598/aeee.v16i4.2773 | cs |
dc.rights | © Vysoká škola báňská - Technická univerzita Ostrava | |
dc.rights | Attribution-NoDerivatives 4.0 International | * |
dc.rights.uri | http://creativecommons.org/licenses/by-nd/4.0/ | * |
dc.subject | ECG | cs |
dc.subject | EDR amplitude algorithm | cs |
dc.subject | miniaturized | cs |
dc.subject | respiratory frequency | cs |
dc.subject | wireless sensor | cs |
dc.title | Optimization of the position of single-lead wireless sensor with low electrodes separation distance for ECG-derived respiration | cs |
dc.type | article | cs |
dc.identifier.doi | 10.15598/aeee.v16i4.2773 | |
dc.rights.access | openAccess | cs |
dc.type.version | publishedVersion | cs |
dc.type.status | Peer-reviewed | cs |