dc.description.abstract | In the article an improved method of measurements synchronization which uses the cross correlation function and
calibration patterns is presented. The previous version of the method turned out to be not precise enough if angle probes were
used during the measurements. Complementing the existing procedures with additional resampling of the signal eliminated in
practice the mentioned problems. The presented calibration strategy reduces the time of preparing multi-channel flow
detector for the test, allows the correction of spatial placement of the probes and allows to improve the repeatability of the
measurements. It also allows to use angle probes and it is suitable for different research methods (echo, shadow or tandem
method). | en |
dc.format.extent | 276036 bytes | cs |
dc.format.mimetype | application/pdf | cs |
dc.language.iso | en | en |
dc.publisher | Žilinská univerzita v Žiline. Elektrotechnická fakulta | en |
dc.relation.ispartofseries | Advances in electrical and electronic engineering | en |
dc.relation.uri | http://advances.utc.sk/index.php/AEEE | en |
dc.rights | Creative Commons Attribution 3.0 Unported (CC BY 3.0) | |
dc.rights | © Žilinská univerzita v Žiline. Elektrotechnická fakulta | |
dc.rights.uri | http://creativecommons.org/licenses/by/3.0/ | |
dc.title | Improving the method of the calibration multichannel ultrasonic flaw detectors for measurements with angle problems | en |
dc.type | article | en |
dc.rights.access | openAccess | |
dc.type.version | publishedVersion | cs |
dc.type.status | Peer-reviewed | cs |