Design and development of a novel Invasive Blood Pressure simulator for patient's monitor testing

dc.contributor.authorBibbo, Daniele
dc.contributor.authorKijonka, Jan
dc.contributor.authorKudrna, Petr
dc.contributor.authorPenhaker, Marek
dc.contributor.authorVávra, Petr
dc.contributor.authorZonča, Pavel
dc.date.accessioned2020-02-26T09:26:29Z
dc.date.available2020-02-26T09:26:29Z
dc.date.issued2020
dc.description.abstractThis paper presents a newly-designed and realized Invasive Blood Pressure (IBP) device for the simulation on patient’s monitors. This device shows improvements and presents extended features with respect to a first prototype presented by the authors and similar systems available in the state-of-the-art. A peculiarity of the presented device is that all implemented features can be customized from the developer and from the point of view of the end user. The realized device has been tested, and its performances in terms of accuracy and of the back-loop measurement of the output for the blood pressure regulation utilization have been described. In particular, an accuracy of ±1 mmHg at 25 °C, on a range from −30 to 300 mmHg, was evaluated under different test conditions. The designed device is an ideal tool for testing IBP modules, for zero setting, and for calibrations. The implemented extended features, like the generation of custom waveforms and the Universal Serial Bus (USB) connectivity, allow use of this device in a wide range of applications, from research to equipment maintenance in clinical environments to educational purposes. Moreover, the presented device represents an innovation, both in terms of technology and methodologies: It allows quick and efficient tests to verify the proper functioning of IBP module of patients’ monitors. With this innovative device, tests can be performed directly in the field and faster procedures can be implemented by the clinical maintenance personnel. This device is an open source project and all materials, hardware, and software are fully available for interested developers or researchers.cs
dc.description.firstpageart. no. 259cs
dc.description.issue1cs
dc.description.sourceWeb of Sciencecs
dc.description.volume20cs
dc.identifier.citationSensors. 2020, vol. 20, issue 1, art. no. 259.cs
dc.identifier.doi10.3390/s20010259
dc.identifier.issn1424-8220
dc.identifier.urihttp://hdl.handle.net/10084/139319
dc.identifier.wos000510493100259
dc.language.isoencs
dc.publisherMDPIcs
dc.relation.ispartofseriesSensorscs
dc.relation.urihttps://doi.org/10.3390/s20010259cs
dc.rights© 2020 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license.cs
dc.rights.accessopenAccesscs
dc.rights.urihttp://creativecommons.org/licenses/by/4.0/cs
dc.subjectinvasive blood pressurecs
dc.subjectmedical devicescs
dc.subjectexciter voltagecs
dc.subjectwaveform simulationcs
dc.titleDesign and development of a novel Invasive Blood Pressure simulator for patient's monitor testingcs
dc.typearticlecs
dc.type.statusPeer-reviewedcs
dc.type.versionpublishedVersioncs

Files

Original bundle

Now showing 1 - 1 out of 1 results
Loading...
Thumbnail Image
Name:
1424-8220-2020v20i1an259.pdf
Size:
5.01 MB
Format:
Adobe Portable Document Format
Description:

License bundle

Now showing 1 - 1 out of 1 results
Loading...
Thumbnail Image
Name:
license.txt
Size:
718 B
Format:
Item-specific license agreed upon to submission
Description: