Analysis of the planar point identification accuracy in CMM measurements

Loading...
Thumbnail Image

Downloads

5

Date issued

Journal Title

Journal ISSN

Volume Title

Publisher

MDPI

Location

Signature

License

Abstract

The paper presents the results of the investigations on the direction-dependent accuracy of the point identification during contact probe measurements with a coordinate measuring machine (CMM). Considering the contact point identified by an orthogonal to the surface probe movement, the transformation of coordinates was made in order to calculate the displacement of the measured point. As a result, the positioning accuracy was estimated in three axes. The experiments demonstrated a strong dependence of the displacement on the declination angle. Moreover, it was found that the directional surface texture which provided different roughness in perpendicular directions, had an impact on the positioning accuracy.

Description

Subject(s)

coordinate measuring machine, precision metrology, accuracy, probing point, error minimization

Citation