Electrical parameters measurement and testing of implantable cardioverters - defibrillators

Loading...
Thumbnail Image

Downloads

8

Date issued

Authors

Penhaker, Marek
Kašík, Vladimír
Korpas, David

Journal Title

Journal ISSN

Volume Title

Publisher

Sigma Not

Location

Není ve fondu ÚK

Signature

Abstract

The aim of the work was measurement and testing of implantable cardioverter - defibrillator (ICDs) device. There was realised electronic measuring circuit which substitutes human body tissue during the device accuracy testing. This circuit enables to measure defibrillation shock discharge in the whole energy range (0.1- 41J). The presented measuring and testing procedures can easy check-up precision of parameters and functionality of embedded electronic device circuit’s part of the ICDs. Presented suggestion and results of this work can be implemented as verification procedure in various types of ICDs and also pacemakers before their implantation as well.

Description

Subject(s)

implantable Cardioverter, defibrillation shock, energy, measurement, programmer, defibrylator, kardiostymulatory

Citation

Przegląd elektrotechniczny. 2011, r. 87, nr. 6, s. 186-189.