Electrical parameters measurement and testing of implantable cardioverters - defibrillators
Loading...
Downloads
8
Date issued
Authors
Penhaker, Marek
Kašík, Vladimír
Korpas, David
Journal Title
Journal ISSN
Volume Title
Publisher
Sigma Not
Location
Není ve fondu ÚK
Signature
Abstract
The aim of the work was measurement and testing of implantable cardioverter - defibrillator (ICDs) device. There was realised electronic measuring circuit which substitutes human body tissue during the device accuracy testing. This circuit enables to measure defibrillation shock discharge in the whole energy range (0.1- 41J). The presented measuring and testing procedures can easy check-up precision of parameters and functionality of embedded electronic device circuit’s part of the ICDs. Presented suggestion and results of this work can be implemented as verification procedure in various types of ICDs and also pacemakers before their implantation as well.
Description
Subject(s)
implantable Cardioverter, defibrillation shock, energy, measurement, programmer, defibrylator, kardiostymulatory
Citation
Przegląd elektrotechniczny. 2011, r. 87, nr. 6, s. 186-189.
Item identifier
Collections
Publikační činnost VŠB-TUO ve Web of Science / Publications of VŠB-TUO in Web of Science
OpenAIRE
Publikační činnost Katedry kybernetiky a biomedicínského inženýrství / Publications of Department of Cybernetics and Biomedical Engineering (450)
Články z časopisů s impakt faktorem / Articles from Impact Factor Journals
OpenAIRE
Publikační činnost Katedry kybernetiky a biomedicínského inženýrství / Publications of Department of Cybernetics and Biomedical Engineering (450)
Články z časopisů s impakt faktorem / Articles from Impact Factor Journals